As the power source for dental clinics, have you conducted maintenance and upkeep on air compressors and suction machines under high temperature conditions?
Q:What maintenance and upkeep is required for the equipment in the Dainuo computer room?
ATTENTION
Before carrying out any maintenance or repairs, please confirm that the air compressor or suction machine is powered off.
Work beyond regular maintenance and repair shall be carried out by designated staff from Dainuo, using only parts specified by the manufacturer.
air compressor
1. Discharge of condensed water:
Condensate water is easily formed during air compression operation. If an air compressor without a drying tower is used, condensate water is prone to accumulate
In the air compressor's air tank, due to its enclosed nature, prolonged exposure to high temperatures can easily lead to the growth of microorganisms and bacteria, causing contamination of the compressed air and also polluting the air tank.
Solution:
Open the machine's drain valve to completely evacuate the water and air from the tank. The drainage cycle can be determined based on the local humidity. If the humidity in the area is low, it can be done once
Discharge once a month, or once a day or once a week if the humidity is too high
Dainuo air compressors can be equipped with various drying towers and automatic drainage devices. The inner tank is coated with a sterilizing and anti-corrosive layer to ensure the purity and dryness of the gas.
2. Setting of air compressor room:
The higher the temperature, the thinner the air becomes (just like how the efficiency of an air compressor decreases in high altitude areas), making it prone to air
The efficiency of the compressor decreases, causing the air compressor to spend more time in the loading state and carry more load, resulting in more heat generated by the air compressor. Generally, the air compressor operates
The operating environment should be below 30℃-40℃. Prolonged exposure to high temperatures can damage the core components of the air compressor, shorten its service life, and cause equipment failure.
Solution:
It is necessary to ensure a good ventilation environment in the computer room. Additional exhaust fans can be installed to blow out hot air and thereby reduce the temperature.
Avoid placing heat sources with high temperatures around the air compressor, as this may cause the machine to inhale air at a higher temperature.
If conditions permit, a fresh air system can be employed to maintain the ideal temperature of the equipment in the computer room at around 25℃.
3. Replace the filter element regularly:
Regularly replacing the filter element of the silencer or filter can extend the service life of the air compressor and also ensure the long-term preservation of high-quality compressed air
The replacement of filter cartridges can prevent infections caused by bacteria or viruses. Especially during the COVID-19 pandemic, it is crucial to ensure that the air used in dentistry is clean and dry.
Suction machine
1. Cleaning of suction tubing:
After the treatment, you can use the strong and weak suction tubes of the dental chair to flush the negative pressure pipeline by inhaling a glass of clean water. Avoid flushing during high temperatures to prevent stains
Bacteria may breed in the residual suction pipeline. It is prohibited to use chlorine-based disinfectants for cleaning, as this may cause damage to the machine in the long run and reduce its service life.
2. Filter cleaning:
It is recommended to remove and flush the equipment filter once a week. This prevents the filter from becoming clogged and ensures that the interior of the machine remains a sealed space for extended periods
High temperatures are conducive to the proliferation of bacteria, which can cause damage to the suction system.
3. Safety operation requirements for negative pressure equipment:
The suction machine needs to be placed in a separate room and physically separated from the air compressor. During operation, the suction machine will generate a large amount of
For thermal energy, if conditions permit, a ventilation system, refrigeration system, and fresh air system can be reserved. The operating temperature should be maintained at a negative pressure and within the range of 5℃-40℃.
4. The equipment exhaust port should be vented to the outside:
According to regulations, negative pressure equipment must exhaust the waste gas to an outdoor safe area, and cannot be discharged into the clinic's air conditioning exhaust system or ceiling area. Each unit
The equipment needs to have an independent exhaust pipe. When the equipment exhausts, the temperature may reach around 60℃-70℃. Take precautions to avoid burns.
For more tips on maintenance and repair, please consult local technical personnel in Dailuo.
Jiangsu | Shanghai | Chengdu
Dainuo Medical Technology specializes in the integrated services of research and development, production, and sales of medical equipment for oral and respiratory oxygen therapy.
With the determination of "ten years to sharpen a sword" and the spirit of craftsmanship, we aim to bring customers a better experience.
Adhering to the core values of "customer first, passion, integrity, and progressiveness", Dainuo Medical Technology strives through continuous research and innovation, and
We continuously pursue high-quality details and are committed to becoming an excellent and reliable comprehensive service provider of medical equipment.
Dainuo Medical Technology [Subscription Account]
Contact us: 400-163-8878
Enhance the global status of China's dental equipment
相关文章
欢迎给我们留言,
我们会尽快回复你!