In the summer of 2022,
temperatures stayed above 40℃ for weeks on end,
putting everyone in an extreme heat mode whenever going out.
As the core power equipment of dental clinics,
have you carried out maintenance and upkeep for air compressors and suction units under high-temperature conditions?

 

Q: What maintenance and upkeep are required for the equipment in the Dental equipment room?

 

1

Attention

English Translation

Before conducting any maintenance or repair work, ensure the air compressor or suction unit is completely powered off.
 
All work beyond daily maintenance and routine upkeep shall be performed by personnel designated by Dainuo, and only original manufacturer-specified parts shall be used.
 

Air Compressor

  1. Condensate Drainage
     
    Condensed water is easily generated during air compression. For air compressors without a drying tower, condensate tends to accumulate inside the air tank. Due to the sealed structure, prolonged exposure to high temperatures will facilitate the growth of microorganisms and bacteria, resulting in contaminated compressed air and internal corrosion of the air tank.
 
Solutions
 
Open the drain valve of the equipment to fully discharge all water and residual air from the tank. The drainage frequency can be adjusted according to local humidity conditions. In low-humidity areas, drain once a month; in high-humidity regions, drain daily or weekly as required.
 
Dainuo air compressors can be equipped with various drying towers and automatic drainage devices. The inner tank is coated with a sterilization and anti-corrosion layer to guarantee the purity and dryness of compressed air.
 
  1. Equipment Room Layout for Air Compressors
     
    The higher the ambient temperature, the thinner the air will become, which is similar to the reduced operating efficiency of air compressors at high altitudes. This will lower the overall working efficiency, force the compressor to stay under loaded operation for longer periods, increase operational load, and generate excessive heat. Generally, the operating ambient temperature for air compressors should be controlled between 30℃ and 40℃. Long-term high-temperature exposure will damage core components, shorten service life and trigger frequent equipment failures.
 
Solutions
 
Ensure good ventilation in the equipment room. Additional exhaust fans can be installed to expel hot air and lower the indoor temperature.
 
Keep high-temperature heat sources away from the air compressor to prevent the equipment from inhaling overheated ambient air.
 
If conditions permit, install a fresh air system to maintain an optimal ambient temperature of around 25℃ for indoor equipment.
 
  1. Regular Filter Replacement
     
    Regular replacement of silencer and filter elements can extend the service life of the air compressor and maintain stable, high-quality compressed air supply in the long run. Timely filter replacement effectively prevents infections caused by bacteria and viruses. Especially amid the COVID-19 pandemic, it is critical to ensure the compressed air used in dental treatments remains clean and dry.
 

Suction Unit

  1. Suction Pipe Cleaning
     
    After each treatment, use the strong and weak suction pipes of the dental chair to flush the negative pressure pipeline by drawing in a cup of clean water. Avoid delayed cleaning in high-temperature environments, as residual waste in pipelines will lead to massive bacterial growth. Chlorine-containing disinfectants are prohibited for cleaning; long-term use will cause irreversible damage to the equipment and shorten its service life.
     
  2. Filter Cleaning
     
    It is recommended to disassemble and rinse the equipment filter on a weekly basis. This prevents filter blockage and maintains a sealed internal operating environment. High-temperature environments accelerate bacterial reproduction, which may lead to malfunction and damage of the suction system.
     
  3. Safety Operation Requirements for Negative Pressure Equipment
     
    Suction units must be placed in an independent room with physical separation from air compressors. A large amount of heat will be generated during operation. Where possible, equip the room with ventilation, refrigeration and fresh air systems. The operating ambient temperature for negative pressure equipment shall be kept within the range of 5℃ to 40℃.
     
  4. Outdoor Ventilation for Equipment Vents
     
    In accordance with relevant regulations, waste gas discharged by negative pressure equipment must be vented to a safe outdoor area, instead of being discharged into the clinic’s air conditioning exhaust system or ceiling space. Each piece of equipment shall be fitted with an independent exhaust pipe. The exhaust temperature can reach approximately 60℃ to 70℃, so protective measures must be taken to avoid scald injuries.
     
For more professional maintenance and repair guidelines, please consult the local technical team of Dental.

1

Jiangsu|Shanghai|Chengdu

Dental Medical Technology specializes in the integrated services of research and development, production and sales for oral and respiratory oxygen therapy medical equipment.
With the determination of "sharpening a sword for ten years" and the craftsman spirit, we deliver a better experience for customers.
Upholding the core values of "Customer First, Passion, Integrity and Enterprising Spirit", Dental Medical Technology keeps forging ahead through continuous research and innovation.
We constantly pursue high-quality details and strive to become an outstanding and reliable comprehensive medical device service provider.
Dental Medical Technology 【Subscribe to Our Account】
Contact Us: 400-163-8878

 

企业微信截图_16587355797213.jpg

Enhance the international standing of China's dental equipment industry

url: https://www.dynamicgroup.cn/en/news/178.html

相关文章

Can't find any content

Welcome to

contact us!

Tel:+86-512-82627666     Email:support@dynamicgroup.cn    

Address:No. 66 Shunxing Road, Lujia Town, Kunshan City, Suzhou City, Jiangsu Province

Copyright © Jiangsu Dynamic Medical Technology Co., Ltd. 苏ICP备2021047747号
online service

Scan code

online service

Hello, I am the online customer service

X